高级检索
当前位置: 首页 > 详情页

Automated evaluation of typical patient-ventilator asynchronies based on lung hysteretic responses

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Intensive Care Unit, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China [2]Department of Mechanical Engineering & Centrefor Bio‑Engineering, University of Canterbury, Christchurch, NewZealand [3]Taicang Yangtze River Delta Research Institute, Suzhou, China
出处:
ISSN:

关键词: Patient–ventilator asynchrony Mechanical ventilation PV loop Hysteretic lung mechanics Hysteresis loop analysis Intensive care unit

摘要:
Patient-ventilator asynchrony is common during mechanical ventilation (MV) in intensive care unit (ICU), leading to worse MV care outcome. Identification of asynchrony is critical for optimizing MV settings to reduce or eliminate asynchrony, whilst current clinical visual inspection of all typical types of asynchronous breaths is difficult and inefficient. Patient asynchronies create a unique pattern of distortions in hysteresis respiratory behaviours presented in pressure-volume (PV) loop.Identification method based on hysteretic lung mechanics and hysteresis loop analysis is proposed to delineate the resulted changes of lung mechanics in PV loop during asynchronous breathing, offering detection of both its incidence and 7 major types. Performance is tested against clinical patient data with comparison to visual inspection conducted by clinical doctors.The identification sensitivity and specificity of 11 patients with 500 breaths for each patient are above 89.5% and 96.8% for all 7 types, respectively. The average sensitivity and specificity across all cases are 94.6% and 99.3%, indicating a very good accuracy. The comparison of statistical analysis between identification and human inspection yields the essential same clinical judgement on patient asynchrony status for each patient, potentially leading to the same clinical decision for setting adjustment.The overall results validate the accuracy and robustness of the identification method for a bedside monitoring, as well as its ability to provide a quantified metric for clinical decision of ventilator setting. Hence, the method shows its potential to assist a more consistent and objective assessment of asynchrony without undermining the efficacy of the current clinical practice.© 2023. BioMed Central Ltd., part of Springer Nature.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院分区:
出版当年[2023]版:
大类 | 4 区 医学
小类 | 4 区 工程:生物医学
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 工程:生物医学
JCR分区:
出版当年[2023]版:
Q3 ENGINEERING, BIOMEDICAL
最新[2023]版:
Q3 ENGINEERING, BIOMEDICAL

影响因子: 最新[2023版] 最新五年平均 出版当年[2023版] 出版当年五年平均 出版前一年[2022版]

第一作者:
第一作者机构: [1]Intensive Care Unit, The Fourth Hospital of Hebei Medical University, Shijiazhuang, China
通讯作者:
通讯机构: [2]Department of Mechanical Engineering & Centrefor Bio‑Engineering, University of Canterbury, Christchurch, NewZealand [3]Taicang Yangtze River Delta Research Institute, Suzhou, China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:39770 今日访问量:0 总访问量:1333 更新日期:2025-05-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 河北医科大学第四医院 技术支持:重庆聚合科技有限公司 地址:河北省石家庄市健康路12号